Milwaukee Instruments Doll MA871 Manual do Utilizador Página 11

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 16
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 10
www.milwaukeeinst.com
11
5. Using plastic pipettes, rinse prism and sample well with distilled or deionized water.
Wipe dry. The instrument is ready for the next sample.
MAKING A STANDARD % BRIX SOLUTION
To make a Brix Solution, follow the procedure below:
Place container (such as a
glass vial or dropper bottle that has a cover) on an
analytical balance.
Tare the balance.
To make an X BRIX solution weigh out X grams of high purity Sucrose (CAS #: 57-50-1)
directly into the container.
Add distilled or deionized water to the container so the total weight of the solution is
100g.
Note: Solutions above 60% Brix need to be vigorously stirred or shaken and heated
in a water bath to roughly 40 °C (104 °F). Remove solution from bath when
sucrose has dissolved. Cool completely before use. The total quantity can be
scaled proportionally for smaller containers but accuracy may be sacrificed.
Example with 25% Brix:
% Brix g Sucrose g Water g Total
25 25.000 75.000 100.000